Traceable® Conductivity/TDS/Salinity Pens Complete Description
Three units provide for conductivity, conductivity/total dissolved solids, and salinity. Pen fulfill requirements for CAP, ASTM, NCCLS, CLSI, ACS, CLIA, AOAC, EPA, APHA, AWWA, WEF, USGS, USP, ISO, and Federal/State regulations.
Easy to use and waterproof
Simply turn on, insert, and simultaneously read result and temperature. Easiest unit ever designed for routine analysis and quality control. Elimination of operator technique permits everyone in the lab to report identical readings. Tough, chemical-resistant, waterproof ABS plastic housing assures a long life in severe lab or harsh plant environments. Large 1/2-inch-high LCD digits are easy to read. Dual display shows reading and temperature. HOLD button freezes the reading on the display.
Calibration using solution standards
Arrow keys allow fast setting to calibration standards. Calibration data is saved even when unit is turned off. Instant-response probe and solid-state thermistor provide precise readings. Temperature compensation permits readings to be automatically referenced to the international standard of 25°C or user may set a programmable temperature coefficient. To assure accuracy an individually serial-numbered Traceable® Certificate is provided from our ISO 17025 calibration laboratory accredited by A2LA. It indicates traceability to standards provided by NIST (National Institute of Standards and Technology). Waterproof unit is 6-1/2 x 1-2/5 x 1-1/4 inches, weight is 2-1/4 ounces. Replacement battery Cat. No. 1039.
Control Company Cat. No. 4365 Traceable® Conductivity Pen
Control Company Cat. No. 4366 Traceable® Conductivity/Total Dissolved Solids Pen
Control Company Cat. No. 4367 Traceable® Salinity Pen
| Specifications Chart |
| Cat. No. |
4365 |
4366 |
4367 |
| Description |
Traceable® Conductivity Meter |
Traceable® Conductivity/
Total Dissolved Solids Meter |
Traceable® Salinity Meter |
| Conductivity Range |
0 to 1999 microsiemens
0 to 19.99 millisiemens |
0 to 1999 microsiemens
0 to 19.99 millisiemens |
– |
| Dissolved Solid Range |
– |
0 to 1999 parts per million
0 to 19.99 parts per thousand |
– |
| Salinity Range (NaCL) |
– |
– |
0.00 to 10.00 ppt 10.1 to 70.0 ppt |
| Resolution |
1 microsiemen |
1 microsiemen/1 part per million |
0.01 ppt |
| Accuracy |
±1% full scale + 1 digit |
±1% full scale + 1 digit |
±2% full scale + 1 digit |
| Temperature Range |
0.0 to 50.0°C 32.0 to 122.0°F |
0.0 to 50.0°C 32.0 to 122.0°F |
0.0 to 50.0°C 32.0 to 122.0°F |
| Temperature Resolution |
0.1° |
0.1° |
0.1° |
| Temperature Accuracy |
±0.5° |
±0.5° |
±0.5° |
| Settable Temp. Coefficient |
0.0 to 4.0% per °C |
0.0 to 4.0% per °C |
– |
| Automatic Temp Compensation |
20° and 25° |
20° and 25° |
25° |
| Total Dissolved Solids Factor |
– |
0.4 to 1.00 |
– |
| Use with Calibration Standards |
400 microsiemens to
19.99 millisiemens |
400 microsiemens to 19.99 millisiemens
400 ppm to 19.99 ppt |
0.20 to 7.00% |
|